Non relevant Ultrasonic Indications can usually be identified as one of the following:
- Electrical Interference
- Interference form the transducer (search unit)
- Interference form the surface of the specimen
- Interference caused by mode conversion of the sound beam
- Interference caused by the shape of the specimen
- Interference caused by material structure
Electrical interference can be caused by improper electrical
connections of sound energy from the interface between the wedge and
test specimen surface.

In Immersion Testing, Air bubbles either on the transducer or
specimen can cause reduced signal amplitude form the back surface and at
times also from the front surface.
Surface Interface
A small amount of surface wave energy is usually transmitted in all directions around a transducer as shown below.
If the transducer is near the edge of a plate, a signal may appear on the CRT.

When inspecting with shear waves, it is possible to detect a surface
discontinuity with the small amount of surface waves generated (see
below).

You can determine if the reflection is from a surface wave by running your finger along the surface in front of the transducer.
If the reflection is due to surface waves, the amplitude of the pipe
will drop on the CRT. When your finger is between the transducer and the
interface producing the signal, surface wave indications may not be
cause for rejection, but they should be evaluated.
Interference can be caused as a result of mode conversion in the test specimen as shown below.
However all of these reflections can be ignored as they will appear on the CRT after the first back reflection.

The shape of the specimen can cause false indications as shown below.

The Ultrasonic operator should always know the configuration of the part so that these false indications can be identified.
To verify a possible false indication, the operator should try to
locate the discontinuity from a different location on the specimen.
A non-relevant indication may occur when using a transducer with a large beam spread as shown below.
This indication is easily identified as it is always behind the first
back reflection of the part and would probably be continuous along the
surface of the part.

In the situation above, a plastic shoe machined to the diameter of the part usually reduces the excessive beam spread.
A large grain size may also cause “Noise” or “Hash” on the CRT
screen. Abnormally large grains may result in total loss of back
reflection. A lower examination frequency may help alleviate this
problem.

During weld inspection, non-relevant indications my result from
reflections from the crown and the root of the weld and possibly from
the heat affected zone.

Any indication on the CRT screen that is unusually consistent can be suspected as being non-relevant.
As is shown above, the reflection from the root and crown of the weld
may appear for the entire length of the weld. If the non-relevant
signal is coming from the crown of the weld, it can often be identified
by placing your finger wet with couplant over the suspected area.
If the sound beam is striking the crown, it will be dampened by your
finger, a surface examination, PT or MT, can be used to reduce the
possibility of a surface crack causing the reflection.